Elettra-Sincrotrone Trieste S.C.p.A. website uses session cookies which are required for users to navigate appropriately and safely. Session cookies created by the Elettra-Sincrotrone Trieste S.C.p.A. website navigation do not affect users' privacy during their browsing experience on our website, as they do not entail processing their personal identification data. Session cookies are not permanently stored and indeed are cancelled when the connection to the Elettra-Sincrotrone Trieste S.C.p.A. website is terminated.
More info
OK

Specifications

 


Beamline Flux 

The experimental flux curves for the 2 storage ring operation energies at Elettra (i.e. 2.0 and 2.4 GeV) normalized to the beam current (300 and 160 mA respectively, typical top-up conditions) are shown below. The output flux was characterized immediately downstream of the refocusing mirror, at 14.3 m from the source, by using 2 ionization chambers behind a 0.3mm thick Beryllium window. The output curves are following closely the expected spectral dependence.

Light Source

Type: Bending Magnet Ring energy: 2.0 GeV Ring Energy: 2.4 GeV
Critical Energy 3210 eV 5590 eV
Source Size (nominally)  sx = 0.33 mm
sy = 0.066 mm
sx = 0.46 mm
sy = 0.07 mm

For more information, please refer to Elettra parameters.

Further details on the optical concept for the X-Ray Fluorescence beam transport system are reported in:

  • Optimisation of a compact optical system for the beamtransport at the x-ray fluorescence beamline at Elettra for experiments with small spots;
    W. Jark, D. Eichert. L. Luehl, A. Gambitta;
    Proceedings of SPIE, Vol. 9207 Advances in X-Ray/EUV Optics and Components IX, pp: 92070G (2014);