Pumping and pressure reading
Pagina 2 di 6
Analysis chamber
Turbomolecular pumpcontrolled by DCU in the right experimental rack on/off or browse parameters together with edit parameter and confirmParameters: 001: heating on/off 012 venting on/off 002: standby (slow speed) on/off 309: actual rotation speed 329: safety bearing wear 358: rotor imbalance (manuals pump controller Automatic venting when at ½ of nominal speed. For venting connect nitrogen gas line to the venting valve. Water- and air-cooled. Water cooling is necessary for bake-out. Scroll pumphas no switch. Disconnect/connect the mains cable (orange).For short periods you can just close the valve on the line below the turbopump and leave the scroll pump on. (manual
controlled by the bottom AML TSP2 controller In the left experimental rack. |
Ionization gaugecontrolled by Prevac MG15 controller in the left experimental rack. ![]() Tap the Emis. button to switch the emission on/off. Tap the pressure value to zoom it in. Tap the pressure unit to change it. The filament can burn at pressures above 10-4 mbar! Never switch on until the turbopump is at full speed and the cold-cathode gauge shows low pressure. (manuals sensor controller
controlled by MaxiGauge controller in the left experimental rack, channel 1 |
Ultima modifica il Giovedì, 19 Ottobre 2023 10:08
on/off
or
browse parameters
off/continuous/degas/timed

switch channels
long press for switch on
long press for switch off