Pumping and pressure reading
Pagina 3 di 6
Preparation chamber - bottom
Turbopumping stationcontaining membrane pump and turbopump.Controlled by DCU in the right experimental rack. on/off or browse parameters together with edit parameter and confirmParameters: 001: heating on/off 002: standby (slow speed) on/off 309: actual rotation speed (manuals station turbo membrane controller Automatic venting when at ½ of nominal speed. For venting connect nitrogen gas line to the venting valve. Water-cooled. Water cooling is necessary for bake-out. Titanium sublimation pumpcontrolled by the top AML TSP2 controller In the left experimental rack.![]() off/continuous/degas/timed(manuals pump controller
|
Cold cathode gaugecontrolled by MaxiGauge controller in the left experimental rack, channel 3 switch channels long press for switch on long press for switch off(manuals sensor controller |