Pagina 2 di 2
PBN sample carriers
	IMPORTANT
	When the annealing on PBN sample holders is done in the preparation chamber or load-lock, the true temperatures are approximately 150°C higher than those in the following tables for the Main chamber where the sample carrieres are in good thermal contact with thy cryostat.
	
	"Old" PBN sample holder
- Manipulator cooled by air flow
| Curret (A) | Voltage (V) | Temperature (°C) | 
| 0.4 | 6 | 100°C | 
| 0.6 | 8 | 150°C | 
| 0.7-0.8 | 10 | 200°C | 
| 1.1 | 13.6 | 300°C | 
| 1.2 | 15 | 327°C | 
| 1.6 | 18.1 | 400°C | 
| 2.1 | 22.4 | 500°C | 
	
	
	"New" PBN sample holder
- Manipulator cooled by air flow
| Curret (A) | Voltage (V) | Temperature (°C) | 
| 0.5 | 7.5 | 100°C | 
| 0.8 | 11.3 | 200°C | 
| 1.3 | 15.3 | 300°C | 
| 1.7 | 18.7 | 400°C | 
| 2.1 | 22.2 | 500°C | 
"All metal" PBN sample holder
- Manipulator cooled by air flow| Curret (A) | Voltage (V) | Temperature (°C) | 
| 1 | 12.2 | 270°C | 
| 1.2 | 14.5 | 327°C | 
| 1.8 | 19.4 | 470°C | 
| 2.0 | 21.4 | 504°C | 
"Elena" PBN sample holder
- Manipulator cooled by air flow| Curret (A) | Voltage (V) | Temperature (°C) | 
| 1.0-1.1 | 9.2 | 256°C | 
| 1.8-1.9 | 14.4 | 420°C | 
Large E-beam sample carrier
Filament radiation annealing, manipulator cooled by air flow
Technical Note: Thoriated W filament, 0.2mm, 13-14 threads| Current (A) | Voltage (V) | Power (W) | Temperature(°C) | 
| 1.4 | 1.8 | 2.5 | 50°C | 
| 1.8 | 2.5 | 4.5 | 100°C | 
| 2.5 | 4.4 | 11 | 210°C | 
| 2.9 | 5.4 | 15.7 | 265°C | 
| 3.0 | 5.7 | 17.1 | 283°C | 
| 3.2 | 6.3 | 20.2 | 300°C | 
| 3.5 | 7.4 | 25.9 | 350°C | 
| 3.6 | 7.5 | 27 | 361°C | 
| 3.8 | 8.3 | 31.5 | 388°C | 
| 3.9 | 8.6 | 33.5 | 402°C | 
Estimated sample temperature vs heater current
				![]()  | 
			
				![]()  | 
		
HV E-beam annealing, manipulator cooled by air flow
Warning 1: HV on thermocouple pins!Warning 2: Operation allowed ontly to the staff.
Warning 3: Filament must be grounded.
Warning 4: Never use PID box to monitor the temperature.
Note: Following data (collected in November 2018) are only approximate, they depend on the filament shape and distance from the Mo plate; the pressure and nature of the used gas.
| Filament (A) | Emission (mA) | 
				Pyrometer (°C) Emis.=16  | 
			
				Temperature (°C) +- 30°C  | 
			Comment | 
| 2.3 | 30 | 420 | 506 | |
| 2.4 | 35 | 550 | 623 | |
| 3.8 | 40-42.3 | 570-600 | 638-665 | pO2=5x10-7 mbar | 
| 3.8 | 51.2 | 660 | 736 | |
| 3.8 | 57.4 | 690 | 760 | pN3B3H6=5x10-7mbar | 
Small E-beam sample carrier
Filament radiation annealing, manipulator cooled by air flow
Technical Note: Thoriated W filament, 0.2mm, 13-14 threads| Current (A) | Voltage (V) | Power (W) | Temperature(°C) | Comment | 
| 1.4 | 1.8 | 2.5 | 102 | After >10 minutes | 
| 3.2 | 6.2 | 20 | 430 +- 30 °C | 
HV E-beam annealing, manipulator cooled by air flow
Warning 1: HV on thermocouple pins!Warning 2: Operation allowed ontly to the staff.
Warning 3: Filament must be grounded.
Warning 4: Never use PID box to monitor the temperature.
Note: December 2018, HV switched on when Filament current is below 1.5A. Pyrometer aimed on the Mo plate close to the crystal sample.
| Filament (A) | Emission (mA) | 
				Pyrometer (°C) Emis.=16  | 
			
				Temperature (°C) +- 30°C  | 
			Comment | 
| 3.7 | 6 | 440°C | 500°C | pO2=2x10-6 mbar | 
| 3.8 | 13.2 | 548°C | 600°C | pO2=2x10-6 mbar | 