Specifications
Specifications
General Properties | |
---|---|
Beamline Energy Resolution | 40 [meV] @ 400 [eV] 180 [meV] @ 900 [eV] |
Beamline Resolving Power | 1 * 104 [E/deltaE] @ 400 [eV] 5 * 103 [E/deltaE] @ 900 [eV] |
Beamline Energy Range | 90 - 1800 [eV] |
Max Flux On Sample | 1 * 1013 [ph/s] @ 200 [eV] 1 * 1012 [ph/s] @ 400 [eV] |
Spot Size On Sample Hor | 100 - 150 [um] |
Spot Size On Sample Vert | 5 - 100 [um] |
Angle Of Incidence Light On Sample Value | 20 - 90 [degrees] |
Control And Data Analysis | |
Control Software Type | LabView based software. |
Data Output Type | Spectra File and Parameter File with Analyzer Settings, Temperature, Pressure, Photon Energy, Reference Current, Sample Polar and Azimuthal Angles, Custom Readings on request. |
Data Output Format | Igor format, ASCII format if requested by the user. |
Softwares For Data Analysis | Commercial Igor software. |
Photon Sources
2 section Kyma Linear Planar Undulator
Type | Undulator |
---|---|
Available Polarization | Linear horizontal |
Source Spectral Brightness | 1 * 1019 [ph/s/mrad^2/mm^2] @ 200 [eV] |
Energy Range | 95 - 1800 [eV] |
Number Of Periods | 98 |
Period | 46 [mm] |
Monochromators
SX700
Energy Range | 85 - 1800 [eV] |
---|---|
Type | Plane Grating Mirror (PGM) |
Resolving Power | 5 * 103 [E/deltaE] @ 900 [eV] 1 * 104 [E/deltaE] @ 400 [eV] |
Number Of Gratings | 2 |
Grating Type | laminar grating - 1220 lines/mm blaze grating - 1220 lines/mm |
Pre-focusing Mirror Type | Cylindrical |
Refocusing Mirror Type | Ellipsoidal |
Endstations or Setup
SuperESCA Main
Spectrometer | Hemispherical Electron Energy Analyzer SPECS Phoibos 150: - mean radius 150 mm - energy resolution < 5 meV - acceptance angle up to ±13° - home made 1D delay line detector | ||||||||
---|---|---|---|---|---|---|---|---|---|
Base Pressure | 8 * 10-11 [mbar] | ||||||||
Detectors Available | 1D Delay Line Detector | ||||||||
Endstation Operative | Yes | ||||||||
Sample | |||||||||
Sample Type | Crystal, Amorphous | ||||||||
Other Sample Type | Nanostructured samples | ||||||||
Mounting Type | Custom mounting plate or modified CTPO sample holder | ||||||||
Required Sample Size | X = 9 [mm], Y = 9 [mm], Z = 1 [mm] | ||||||||
Techniques usage | |||||||||
Absorption / NEXAFS | Absorption spectra are measured in Auger yield. By changing sample orientation, the angle between the electric field E of the polarized light and and the normal to the sample surface can be varied in the range from 20 to 90 degrees. | ||||||||
Photoelectron emission / Angular Resolved PES | The technique is available if the modified CTPO manipulator is used, which allows software controlled angular movements of the samples. | ||||||||
Photoelectron emission / Photoelectron diffraction | Core level intensity is measured as a function of polar and azimuthal emission angles. The technique is available if the modified CTPO manipulator is used, which allows software controlled angular movements of the samples. | ||||||||
Photoelectron emission / Time-resolved studies | Fast XPS allows monitoring the evolution of core level spectra (typical acquisition time between 4 and 10 seconds per spectrum), e.g. while exposing the sample to gases or during an annealing. | ||||||||
Manipulator or Sample stage | |||||||||
Manipulator |
| ||||||||
Sample Holders | |||||||||
Custom sample holder |
| ||||||||
Modified CTPO holder |
|
Detectors
1D Delay Line Detector
Type | Microchannel Plate based delay line detector. |
---|---|
Description | In house developed 1 dimensional delay line detector. The detector is mounted on the SPECS Phoibos 150 electron spectrometer. |
Pixel Size | X = 100 [um] |
Array Size | X = 800 [pixel] |
Passive or Active (Electronics) | Active |
Output Readout Software | LabView based home developed software |
Detection | |
Detected Particle | Electron |
Ultima modifica il Venerdì, 08 Marzo 2019 11:20