Elettra-Sincrotrone Trieste S.C.p.A. website uses session cookies which are required for users to navigate appropriately and safely. Session cookies created by the Elettra-Sincrotrone Trieste S.C.p.A. website navigation do not affect users' privacy during their browsing experience on our website, as they do not entail processing their personal identification data. Session cookies are not permanently stored and indeed are cancelled when the connection to the Elettra-Sincrotrone Trieste S.C.p.A. website is terminated.
More info
OK

Specifications

Accessible SAXS and WAXS range


Small angle:

Real space: 1-140 nm (8 keV)
                     1-63 nm (16 keV)



Wide angle:

angular region from 17 - 85o  (2Theta)



SWAXS:

Simultaneous SAXS- and WAXS-measurements can be performed using a Pilatus3 1M detector for the SAXS signal, and a Pilatus 100k detector for the WAXS-range.

For transmission experiments, also a specially designed vacuum chamber (SWAXS-nose, see Annual Report of 1996/97, p. 32) can be used which allows to use both scattering areas below (for SAXS) and above (for WAXS) the direct beam, respectively. The overall length of the SWAXS-nose in horizontal direction, measured from the sample position, is 512 mm and the fixed sample to WAXS-detector distance is 324 mm. Therefore, at the shortest SAXS camera-length an overlap in the d-spacings covered by the SAXS- and WAXS-detectors, respectively, is possible: then, the common regime lies around 9 Ã… (at 8 keV, using for WAXS the 2Theta = 9.4 - 27.6o range).

Last Updated on Monday, 29 April 2019 15:20