Exposing the sample with gases
Exposing the sample with gases
- Check that LEED, electron analyzer, X-ray source, evaporators, titanium sublimation pump etc. are off and that all gate valves to other chambers (also the electropneumatic valve 10 towards the beamline) are properly closed.
- For the desired exposure in langmuirs (1 L = 10-6 torr×s) calculate the necessary pressure and time. Reasonable times are between 30 s and 30 min. The pressure must not exceed 1×10-6 mbar. This limit is dependent on the gas and exposure time, the rule is that the pressure after the exposure should drop within reasonable time. So be careful especially with water, methanol and similar compounds.
-
On the ionization gauge touchscreen tap the current pressure unit (mbar) and change it to torr if needed:
- Start the Pressure Display software if needed. Display the experimental (ioniz.g.) channel.
- Open slowly the corresponding leak valve. When you reach the desired pressure start the timer.
- Observe and adjust the leak valve if necessary.
- When finished close the leak valve without forcing it.
- Wait for the pressure in the chamber to recover. The analyzer and the X-ray source cannot be operated before the cold-cathode pressure reading is below 5×10-8 mbar due to interlocks.
- Set the ionization gauge controller units back to mbars if needed.
Installing gas bottles (CO, O2, H2...)
|
Installing liquids (H2O, CH3OH...)
|
Quadrupole mass spectrometer
The QMS is used for checking the residual gas composition, possible leaks and purity of the gases. The software Quadstar is described here.- Check that the pressure is below 1×10-6 mbar.
- Connect the QMS power supply to electric network. Two green LEDs will light on the QMS electronics.
-
Open the Quadstar Measure software:
-
Switch on the Emission and multiplier (SEM) voltage in the Setup - SEM/Emission Control menu:
-
Start the Analog scan - SCAC.SAP in the Scan menu:
-
Adjust the parameters using Parameters - Setup and Parameters - Channel menu (mass range, intensity range):
-
Observe the spectrum and save it using the File - Save cycle data into a .sac file if needed:
- Close the scan window.
-
Switch off the Emission and multiplier (SEM) voltage in the Setup - SEM/Emission Control menu:
- Close Quadstar Measure.
- Disconnect the QMS power supply from electricity.
- If you saved the data into .sac file use the Quadstar Dispsav software to Open, display the data and Convert them to ASCII (.asc) files that can be imported into common spreadsheet software like Excel, Origin or Igor for further processing.