Operating the beamline
Page 1 of 4
General
The user-controlled electropneumatic valves of the beamline (labelled by numbers 1-10 in the screenshot adapted from the Beamline Control System (BCS) webpage above). Before using the beam for data acquisition the beamline must be open for at least 30 minutes in order to reach thermal stability of its optical elements (warm-up), mainly of the prefocusing mirror, which is being hit by the most intense light (40 W incident, of which 90% absorbed). The valves can be operated using- the Beamline software, or
- the BCS webpage (single click on the valve) on the MSB-MASTER or MSB-STATION computers, or
- the BCS section of the Kleopatra software (double click on the valve) from the MSB-MASTER computer (or its VNC copy on the MSB-STATION computer).
The other two options open a webpage where the desired command (1) open or (0) close can be selected and Submitted. It will open another webpage with the result of the operation. The status of each valve is indicated by color (green - closed, red - open). The displayed status is refreshed only every 10-40 s so the change might not be immediately visible..
Opening the beamline using the Beamline software
|
Closing the beamline
|
Opening the beamline using the web-based method
|
Closing the beamline
|
Notes
Valves a, b and c cannot be controlled by the user.- Valves a and b are closed by the machine operators before the injection (e.g. in the case of a beam dump) and open after it. If the operators forget to open these valves after the injection call the Control Room (telephone number is on the machine status webpage) and ask kindly to open the machine shutter of the beamline 6.1 Materials Science.
- Valve c is a security fast valve closed automatically if vacuum is broken.
Valve 1 also automatically closes when there is not enough flow of the cooling water of the beamline prefocusing mirror, entrance slit and/or monochromator. Check Water status being OK in the Beamline software or on the web and increase it if needed.
The correct opening/closing order of valves 1-7 is important. The scope is to avoid hitting any of the valves 2-7 by the photon beam when they are closed. Therefore the valve 1 should be open as the last one and closed as the first one, as it is the only one that can support the heat load of the beam (it is water-cooled). Although there are some interlocks that should avoid user errors please respect the correct order.
Open the valve 10 to the analysis chamber only for the period of photoemission measurements and close it immediately after. Double-check that it is closed especially when performing operations leading to pressure increase in the analysis chamber, typically sample transfer, ion sputtering or gas exposure, or during photoemission with X-ray source (to avoid mixing two different excitations). Moreover, some sensitive samples might be damaged by long exposures to the photon beam so there is no reason to illuminate them when not needed.
If the BCS webpage does not open at all and the BCS section of Kleopatra is of grey color and does not react on double-clicks on valves wait for several minutes. If the problem persists follow Troubleshooting - BCS communication problem.
- Prev
- Next >>
Last Updated on Thursday, 09 March 2023 10:29