The photons dispersion system is based on the Padmore VASGM (Variable Angle Spherical Grating Monochromator) scheme and it includes four different interchangeable spherical gratings. The first three gratings guarantee resolving powers of 20000-6000, 20000-6000 and 15000-5000 in the energy ranges 35-200 eV, 200-500 eV and 500-1600 eV respectively.
MONOCHROMATOR: ENERGY RANGES
Grating | Optimized Energy Range | Resolving Power | Flux with slits at 10 mm (high resolution) |
SG1 | 44 eV-349 eV (35 eV-42 eV available under special request) | high resolving power: 20000-6000 | flux>1012 ph/s |
SG2 | 158 eV-597 eV | high resolving power: 20000-6000 | flux>1010 ph/s |
SG3 | 491 eV-1600 eV | high resolving power: 15000-5000 | flux>1010 ph/s |
SG4 | 301 eV-1600 eV | medium resolving power: 10000-2000 |
flux>1011 ph/s |