Twin anode X-ray source

Sample Transfer and Measurement Positions

  1.  Ensure the X-ray source is off and pulled out to the safe position marked by the top red line on the linear transfer.

X-ray source position for
sample transfer
SampleTransferPosition

  1. Transfer the sample.
  2. Move the manipulator to the default position for XPS:

Default XPS position
Theta angle 240°
x 25.0
y 28.0
z 10.0
  1.  Verify sample position in camera view. Orange circle indicates analyzed area (~7 mm).

Analyzed area
AnalyzesSpot



  1. Adjust Z-axis (vertical) position.
  2. Adjust horizontal position.
    1. X and Y must be moved almost simultaneosly. Change each single axis by no more than 1 mm.
    2. Per 1 mm of horizontal asjustment, move Y by 0.87 mm (sin(60°)) and X by -0.5 mm (-cos(60°)).
X,Y positions
Δd (mm) X (mm) Y (mm)   Δd (mm) X Y
0 25.0 28.0 Default position 0 25.0 28.0
-1 24.5 28.9   1 25.5 27.1
-2 24.0 29.7   2 26.0 26.3
-3 23.5 30.6   3 26.5 25.4
-4 23.0 31.5   4 27.0 24.5
-5 22.5 32.3   5 27.5 23.7

  1. Lower X-ray gun to the red line on the linear transfer, approximately 8 mm from the bottom limit. (8 mm equals the width of wrench 7, which is often somewhere around.)

X-ray gun operation position
XGunWorkingPosition
  1. Cover windows 14, 15, 16, and 22 with lead shields.

Lead shields
LeadShields


Start-Up Procedure

  1. Verify the main chamber base pressure is below 5x10-9 mbar.
  2. Turn on the X-ray source power supply and wait at least 10 minutes for electronics thermal stabilization.
  3. Open the water cooling circuit and set the water pressure to aprox. 4.8 bars.
  • Open drain valve first.
  • Slowly open the source valve until the pressure reaches approximately 4.8 bars.
Water cooling
WaterCircuit

  1. Verify all interlock LEDs are green.
Power supply, Interlocks
HVpowerSupply

  1. Start Avantage software.
Avantage
  1. Navigate to the "Instrument" tab (A). The Xray Gun Control panel will appear on the top.
    1. If the panel displays "Device OK" (B), software control is ready.
    2. Expand the control panel by clicking the arrow on its right edge.
X ray Gun Control Panel
ControlPanel
  1. Select 15W from the Spot/Power list in the X-Ray Gun Control panel (C).
  2. Monitor the main chamber pressure.
  3. Press the Apply button (D). The gun ramping up procedure will start. Wait for the "Device OK" (B) message and monitor the pressure, which should remain below 5x10-9 mbar. (Initial pressure spikes up to 1x10-8 mbar mbar are normal.)
  4. Once the pressure stabilizes (decreases), gradually increase the power to 100W, 200W, and finally 300W.
  5. Wait at least 10 minutes for X-ray intensity to stabilize.


Shutdown Procedure

  1. In the X-Ray Gun Control panel, press the Shutdown button and wait for the "Device OK" message to reappear.
  2. Verify that the red "HV ON" indicator on the power supply is off.
  3. Maintain water cooling for at least 10 minutes.
  4. Close the water cooling circuit.
    1. Slowly close the source valve first..
    2. Then, close the drain valve.
  5. Pull out the X-ray gun. Raise the linear transfer until the top red line mark is visible.
  6. Exit Avantage software.

 


Binding energy scale calibration


Pass energy 100 eV, A21 mode:
Convert kinetic energies to binding energies by subtracting the kinetic energy from the Fermi edge positions listed in the following table.


Anode Photon energy (eV) Fermi edge KE (eV) Resolution (eV)
Mg Kα 1253.6 1248.67 0.85
Al Kα 1486.6 1481.72 1.0



Troubleshooting

Under construction.