Venting, bake-out and degassing
Page 3 of 3
Degassing
Some equipment must be properly degassed before use. It can be done at any time but it is convenient to do it until the chamber is still warm after the bake-out, so the degassed molecules adsorb less on other surfaces. Only the electron analyzer detector must be degassed after everything cools completely down!The typical degas procedure consists of step-wise reaching the operating parameters while observing that the pressure does not rise much.
Titanium sublimation pumpThe degas procedure is fully automated by the corresponding controller (degas knob) and consists of slow reaching 25 A filament current and keeping it for 15 min. It is enough to degas one filament. Nevertheless, the pressure significantly rises during every successive sublimation cycle (38.5 A, 1 min) up to 10-8-10-7 mbar. Observe the pressure during first several sublimation cycles after the bake-out and abort the sublimation if the pressure is too high..Estimated time: 30 min. Quadrupole mass spectrometerNo special degas procedure is required. Switch on the filament as described here and keep it on for about 30 min. You can also record the mass spectrum. If there is no leak in the chamber the intensity of mass 14 should be lower than that of mass 15.Estimated time: 30 min Sputter ion gunThe corresponding power supply provides an automated degas procedure activated by the DEGAS knob. It is enough to degas one filament.For details see the corresponding manual. Estimated time: 10 min. LEEDThe filament can be degassed either manually at ~2.50 A or using a built-in automatic procedure for few hours until the beam current is stable and in μA range.For details see the corresponding manual. Estimated time: 3 hours. Knudsen-cell evaporatorsThe evaporators are degassed on first use by reaching the desired parameters.Estimated time: 30 min. |
X-ray sourceThe power supply provides semi-automated procedures for:
Estimated time: 15 hours. Tectra e-beam evaporators
|
Oxford e-beam evaporator
Sample manipulator and sampleThe degas happens at first several cycles of sample annealing.Estimated time: 30 min. Electron analyzer detectorWarning: this is the only degas procedure that must be done after complete cooling down. Wait at least 72 hours after the end of the bake-out!The detector is degassed by slowly increasing the voltage from 0 up to ~2000 V with all excitation sources off. It is done by using the detector voltage scan saved in the yymmdd_start-up.xml file in the current dara folder (D:\Data\20**\inhouse\). For parameters you can use any recent file. For details see the corresponding manual. Estimated time: 30 min. |
- << Prev
- Next
Last Updated on Tuesday, 09 April 2024 11:48