Specifications
Page 4 of 5
Sample stage
Sample manipulator with
X, Y and
Z linear movements,
polar and
azimuth rotations.
Sample heating |
indirect resistive |
Tmax = 1300 K |
direct resistive through the sample |
Tmax = 1300 K |
Sample cooling |
LN2 manipulator |
Tmin = 100 K |
The sample temperature is read by a K-type thermocouple (chromel/alumel), which is mounted on the rear side of the sample, or with an optical pyrometer.
For details see
Sample mounting.