Elettra-Sincrotrone Trieste S.C.p.A. website uses session cookies which are required for users to navigate appropriately and safely. Session cookies created by the Elettra-Sincrotrone Trieste S.C.p.A. website navigation do not affect users' privacy during their browsing experience on our website, as they do not entail processing their personal identification data. Session cookies are not permanently stored and indeed are cancelled when the connection to the Elettra-Sincrotrone Trieste S.C.p.A. website is terminated.
More info
OK

Scienta R3000: non-linear intensity scale correction

R3000 XPS Intensity scale calibration, Angular Modes, BACH, May 2015

Note: All calibration curves were done with detector area set to 640x384 px2, where 640px corresponds to full x (energy) scale, 384px to Y-(angular) scale between 53 and 436px.

Fixed Mode (Calibration → Voltages...→ Area)





Swept Mode




Graphs in higher resolution.

Approximate relation between intensity in swept mode and fixed mode (full detector range):

SES, Igor Pro:
Cps = C / n/ StepTime                   (1)
where,
Cps = couts per second in fixed mode ( approximately what we would see in Calibration → Voltages →Area)
C= counts in spectrum point acquired in swept angular mode
n = number of sweeps

KolXPD , if Io reading is ON and txt file was imported. If ibw file was opened accidently, use formula 1):
Cps = C * Io *N                                (2)
N = Number of steps = 0.124*PassEnergy/EnergyStep