Scienta R3000: non-linear intensity scale
R3000 XPS Intensity scale calibration
Note: All calibration curves were done with detector area set to 640x384 px2, where 640px corresponds to full x (energy) scale, 384px to Y-(angular) scale between 53 and 436px.
Fixed Mode (Calibration → Voltages...→ Area)


Swept Mode


Graphs in higher resolution.
Approximate relation between intensity in swept and fixed mode (full detector range):
SES, Igor Pro:
Cps = C / n/ StepTime
where,
Cps = couts per second in fixed mode ( approximately what we see in Calibration → Voltages →Area)
C= counts acquired in swept angular mode
n = number of sweeps
KolXPD has a function to correct detector non-linearity once the calibration file is provided
- Make sure the spectra are measured in swept, angular, angle-integrated mode and saved as txt files.
- Set following Load options, keep the line with the calibration file path empty in this step.
- Open spectrum in KolXPD
- Multiply the spectrum by the Io value. (If Io signal was recorded, the normalization to Io must be turned back before appying the non-linearity calibration curve).

- Go to Operations → Correction for non-linearity - calibration wave
- Select file with the calibration data (IntensityCalibration2022.txt or IntensityCalibration2015.txt for older data), then click OK.
- Normalize to Io (Divede by the Io value).