Scienta R3000: non-linear intensity scale
Page 2 of 2
KolXPD has a function to correct detector non-linearity once the calibration file is provided
- Make sure the spectra are measured in swept, angular, angle-integrated mode and saved as txt files.
- Set following Load options, keep the line with the calibration file path empty in this step.
- Open spectrum in KolXPD
- Multiply the spectrum by the Io value. (If Io signal was recorded, the normalization to Io must be turned back before appying the non-linearity calibration curve).

- Go to Operations → Correction for non-linearity - calibration wave
- Select file with the calibration data (IntensityCalibration2022.txt or IntensityCalibration2015.txt for older data), then click OK.
- Normalize to Io (Divede by the Io value).