Pumping and pressure reading
Page 2 of 6
Analysis chamber
Turbomolecular pumpcontrolled by DCU in the right experimental rackon/off or browse parameters together with edit parameter and confirm Parameters: 001: heating on/off 012 venting on/off 002: standby (slow speed) on/off 309: actual rotation speed 329: safety bearing wear 358: rotor imbalance (manuals pump controller ) Automatic venting when at ½ of nominal speed. For venting connect nitrogen gas line to the venting valve. Water- and air-cooled. Water cooling is necessary for bake-out. Scroll pumphas no switch. Disconnect/connect the mains cable (orange).For short periods you can just close the valve on the line below the turbopump and leave the scroll pump on. (manual )
controlled by the bottom AML TSP2 controller In the left experimental rack. |
Ionization gaugecontrolled by Prevac MG15 controller in the left experimental rack.Tap the Emis. button to switch the emission on/off. Tap the pressure value to zoom it in. Tap the pressure unit to change it. The filament can burn at pressures above 10-4 mbar! Never switch on until the turbopump is at full speed and the cold-cathode gauge shows low pressure. (manuals sensor controller )
controlled by MaxiGauge controller in the left experimental rack, channel 1 |