Elettra-Sincrotrone Trieste S.C.p.A. website uses session cookies which are required for users to navigate appropriately and safely. Session cookies created by the Elettra-Sincrotrone Trieste S.C.p.A. website navigation do not affect users' privacy during their browsing experience on our website, as they do not entail processing their personal identification data. Session cookies are not permanently stored and indeed are cancelled when the connection to the Elettra-Sincrotrone Trieste S.C.p.A. website is terminated.
More info
OK

Pumping and pressure reading

Load lock

Turbopumping station

containing membrane pump and turbopump.
Controlled by DCU in the right experimental rack.
   on/off
  or browse parameters
  together with edit parameter and confirm
  Parameters:
  001: heating on/off
  002: standby (slow speed) on/off
  309: actual rotation speed
  (manuals station turbo membrane controller )
Automatic venting when at ½ of nominal speed.
For venting connect nitrogen gas line to the venting valve.
Water- and air-cooled. Water cooling is necessary for bake-out.

Full range gauge

controlled by MaxiGauge controller in the left experimental rack, channel 4
   switch channels
  long press for switch on
  long press for switch off
 
(manuals sensor controller )